10 Troubleshooting
Error Possible
Reason
Correction
Connection
Check connection between sensor cable and termi-
nal strip. Check connection boxes, connections for
sensor cable extension for correct connection or
moisture.
Check sensor installation for correct installation and
adjustment / alignment.
Sensor
Check sufficient signal strength (I/O Menu). Apply
coupling paste to the sensors if required.
Transmitter
Call up error memory. J Depending on error mes-
sages take the respective measures (check cables,
plug and socket connections, sensor installation).
f >CPU Error< is indicated consult NIVUS service
personnel.
Investigate the moment of failure in menu RUN –
Trend.
Negative flow direction
Sensors shifted during connection?
Î
Check individual path velocities.
Æ
exchange sensor connections of the respective
path
No indication of flow
(>0< or. >----<)
Programming
Completely check the transmitter parameter set-
tings.
Connection
Check power supply connection.
Check power supply voltage.
Check switch position on connection board.
Power supply
Compare power supply (AC or DC) with transmitter
type.
Unauthorised manufacturer
Æ
use NIVUS memory
card.
Improper memory size?
Æ
use card with permissible
maximum capacity
No screen (black /
flickering)
Memory card
Card formatted on PC?
Æ
send card to NIVUS.
Connection
Check cable connection. Wiring on terminal strip
switched? Cables firmly connected to plugs (re-
tighten screws, pull at cable ends)? Insulation of
single wires unintentionally clamped in?
Display Error
Sensor >X<
Communication
Communication to DSP sensor evaluation disturbed.
Can be checked by pressing the I-key. DSP Sensor
should be indicated in the third line of the following
screen. Contact manufacturer service.