Sensor Installation and connection
6.3.1 Sensor installation
The inserted sensors have to be fastened hard and tight. Use only non-corrosive
fastening material! Before installing the sensor, please check material and wall
thickness of the pipe. The material should be cast steel, steel or stainless steel.
These materials can be selected directly from the start-up menu.
Before mounting the sensors, the transmitter must be programmed (see chap-
ter 8).
Subsequently install the first sensor in flow direction according to the respective
marking.
Apply coupling paste to the surface touching the pipe, thread the clamp band
and attach the sensors on the pipe with the arrow pointing in flow direction (ac-
cording to path arrangement).
Fig. 6-3
Sensor pair (sensor A and B)
Please use the delivered clamp bands (mounting accessory) for mounting. First
shorten the clamp band to the length required. Then insert the clamp band into
the slotted clamp and bend it over (see Fig. 6-4). Please observe to have the
marked side of the slotted clamp facing upwards.
Fig. 6-4
Clamp band for clamp on sensors
Now put on the turnbuckle, fix it by folding over and tighten the knurled screw
using an Allen
®
key.
Please observe to have a sufficient amount of coupling paste being applied to
the sensor face.
Attach the second sensor parallel to the pipeline. The transmitter indicates the
required distance to the previously mounted sensor. Then install the sensor as
described before.