Instruction manual
NivuSonic Clamp on
®
The internal memory has a capacity of 90 days. Subsequently, saved data will
be overwritten erasing the oldest data sets first.
Indicated measurement values will be scaled automatically during scrolling and
therefore might change in order to ensure the best reading available.
The memory time interval can be set in the PAR - Memory Mode - Interval
menu. The factory default setting is set to an interval of 2 minutes.
1 Graphic
2 Minimum value
3 Maximum value
4 Indicated period
Fig. 8-5
Trend graphic example
All trend graphic values saved previously will get lost if the memory interval or
another parameter is going to be modified.
8.4
Display Menu (EXTRA)
This menu allows to modify settings such as basic screen, units, language as
well as the display itself. The following submenus are available:
Fig. 8-6
Extra submenus
Fig. 8-7
Unit system selection
NivuSonic CO - Rev. 00 as of 08.12.2010
Page 51