Instruction manual
NivuSonic Clamp on
®
12 Maintenance and Cleaning
The device Type NivuSonic CO is designed to be virtually maintenance-free and
free of material wear and does not need to be calibrated.
If required clean the transmitter enclosure if with a dry, lint-free cloth. For heavy
pollution NIVUS recommends the use of surface-active agents.
The use of abrasive cleansing agents is not allowed.
If you wish to clean the enclosure surface with a damp cloth please disconnect
the unit from mains before.
No hard objects such as wire brushes, rods, scrapers or similar shall be used to
clean the sensors.
Never use scrubbing, scouring cleansing agents or other tools / media which
might have abrasive effects or organic solvents to clean sensor cables and sen-
sors.
Using a high pressure cleaner may lead to measurement failures and thus is not
allowed.
The use of high pressure cleaners may damage the sensors and hence may lead
to measurement failures. Therefore it is not allowed to use such cleaners.
The measurement place must be recalibrated (see chapter 7) after the sensors
have been removed from the holders for cleaning purposes.
13 Emergency
In case of emergency
-
press the emergency-off button of the main system or
-
set the slide switch (see Fig. 6-15) a on the unit to OFF.
14 Dismantling/Disposal
The device shall be disposed according to the local regulations for electronic
products.
NivuSonic CO - Rev. 00 as of 08.12.2010