Instruction manual
NivuSonic Clamp on
®
8
Parameter Setting
8.1
Quick Guide Parameter Setting (Quick Start )
In case of standard applications – part filled standard channel; level and flow ve-
locity measurement by one pair of sensors; 1x mA output for flow rate; 1x im-
pulse output - normally just a few basic settings, such as the ones mentioned
below are required:
1. Install and connect transmitter and sensor as described in chapter 6
2. Connect power supply
3. Menu: EXTRA – Units: select units for flow (l/s), velocity (m/s), level/height
(m) and total (m³). (units in brackets = default settings)
4. Menu: PAR – Measurement place – Channel profile: select profile
5. Menu: PAR – Measurement place – Channel dimensions: enter channel di-
mensions
6. Menu: PAR – LDV Parameter – path arrangement: enter number of paths
(enter number of sensor pairs) and select operating mode (v-echo).
Additional Settings
7. Menu: EXTRA – Display: adjust brightness and contrast if necessary
8. Menu: EXTRA – System time: adjust time if necessary
9. Menu: PAR – Measurement place – Name of measurement place: enter the
name of the measurement place
10. Menu: PAR – analog outputs – Function: enable analog output 1 (flow)
11. Menu: PAR – analog outputs – output span: select output span
12. Menu: PAR – analog outputs – measurement span: select measurement
span
13. Menu: PAR – analog outputs – error mode: define what level the analog
output should take in case of error
14. Menu: PAR – Relay outputs – Function: enable relay 1 (select pos-total im-
pulse)
15. Menu: PAR – Relay outputs – Pulse parameter: set impulse value and dura-
tion
16. Exit parameter setting. Save values by entering code number 2718 a
NivuSonic CO - Rev. 00 as of 08.12.2010