Instruction manual
NivuSonic Clamp on
®
Channel number
Select from analog outputs 1–4 which one is to be set by using the following
parameters.
Name
No entry required. Setting a name here is helpful only if the analog output is to
be saved on memory card. This name will be saved on the storage medium ex-
clusively. The procedure is the same as described in >PAR/Measurement
Place/Name of Measurement Place<.
Function
Functions are going to be assigned to the analog output which has been chosen
by entering the >Channel Number<.
Select:
-
inactive (no analog signal output)
-
Flow rate output (output of analog signal which is proportional to calculated
flow volume)
-
Velocity output (output of analog signal which is proportional to mean flow
velocity averaged from measured individual velocities)
-
Temperature water (output of measured water temperature as analog signal)
Fig. 8-29
Selection of analog output functions
Output range
If desired toggle between measurement ranges of 0-20 mA or 4-20 mA here.
Measurement span
Define the span of the enabled analog output here.
Negative values
can be
entered as well!
Fig. 8-30
Measurement span
Example:
A measurement place is partially tending to backwater formation. Negative val-
ues shall be recorded as well, the following recording or process conducting sys-
tem however has only one analog input left available. In this case the analog
output signal is set to have a "floating” behaviour.
NivuSonic CO - Rev. 00 as of 08.12.2010
Page 61