Instruction manual
NivuSonic Clamp on
typedef struct scalar {
float
min;
float
max;
}
The data type "float" corresponds to the "IEEE-Single"-Definition, transmission is
executed in "big-endian".
Array member "min" represents the corresponding value for 0 digits, "max"
represents the measurement value at 32767
digits.
Verification of registers
The registers "Read Input Register" and "Read Holding Register" can be verified
directly on the unit under
I/O > Communication > Modbus > Read Input Reg.
(or Read Holding Reg.)
Pressing the "ALT" key allows to toggle the screen from decimal to hexadecimal
representation.