Instruction manual
NivuSonic Clamp on
®
-
49:
Status bits to 33-48
-
50-65: Sensor2 Gate 1-16 (scaled)
-
66:
Status bits to 50-65
-
67-82: Sensor3 Gate 1-16 (scaled)
-
83:
Status bits to 50-65
-
84-99: Sensor4 Gate 1-16 (scaled)
-
100:
Status bits to 84-99
-
101-108 average velocities V1-4 or (path 1-8 (scaled))
-
109
Status bits to 101-108
Example:
Address 1
Flow [m³/s]
Scaled from –32767 to +32767 corresponds to 1 Digit = 1 l/s
Address 2
Level [m]
Scaled from –32767 to +32767 corresponds to 1 Digit = 1 mm
Address 3
Velocity [m/s]
Scaled from –32767 to +32767 corresponds to 1 Digit = 1 mm/s
Address 4
Temperature [°C]
Scaled from –32767 to +32767 corresponds to 1 Digit = 0.01 °C
Addresses 7 – 14
Analog Inputs [%]
Scaled from –10000 to +10000 corresponds to 1 Digit = 0.01 %
Addresses 15 – 22
Analog Outputs [%]
Scaled from –10000 to +10000 corresponds to 1 Digit = 0.01 %
The status registers (addresses 6, 23, 32, 49, 66, 83, 100) indicate the validity of
the according measurement values. The value is invalid as soon as the bit is set
at the relevant place.
NivuSonic CO - Rev. 00 as of 08.12.2010
Page 105