4.1 Overview of measurement and processing
FRA5097
4-5
4.1.4 Display mode
In the display mode setting, representation methods of graphic display for analyzed data in the
LCD screen, i.e., gain (or amplitude) vs. phase, are specified/set.
The display mode is set through the menu [Graph][FORMAT][DISPLAY MODE]. The data
displayed in the graph depend on the analysis mode and display units setting.
The following explains the display mode when the display unit is set to GAIN. For the display
mode when the display unit is set to IMPEDANCE, see 5. Impedance display function.
Combinations other than those shown in Table 4-3 List of display modes are invalid. When an
invalid combination is specified, no graph is displayed, or the previously shown graph is deleted.
Since the data is retained in this case, a graph is displayed when a valid combination is specified.
Table 4-3 List of display modes (1) for analysis mode: (CH1) or (CH2)
[DISPLAY MODE]
Display Item
X axis Y1 axis Y2 axis
X axis
Y1 axis
Y2 axis
Remark
logF
F
dBR
logR
R
θ
-
Frequency
Amplitude Phase/none
logF
F
θ
-
Frequency
Phase
logF
F
logA
log(-A)
logB
log(-B) Frequency
Real amplitude
Imaginary
amplitude
logF
F
A B
Frequency
Real amplitude
Imaginary
amplitude
Bode chart
θ
dBR
logR
R
-
Phase
Amplitude
Nichols chart
A B
-
Real
amplitude
Imaginary
amplitude
Nyquist diagram
A -B
-
Real
amplitude
Negative
imaginary
amplitude
Cole cole plot
Table 4-3 List of display modes (2) for analysis mode: (CH1/CH2) or (CH2/CH1)
[DISPLAY MODE]
Display Item
X axis Y1 axis Y2 axis
X axis
Y1 axis
Y2 axis
Remark
logF
F
dBR
logR
R
θ
-
Frequency
Gain
Phase/none
logF
F
θ
-
Frequency
Phase
logF
F
logA
log(-A)
logB
log(-B) Frequency
Real gain
Imaginary
gain
logF
F
A
B
Frequency
Real gain
Imaginary
gain
Bode chart
θ
dBR
logR
R
-
Phase
Gain
Nichols chart
A B
-
Real gain
Imaginary gain
Nyquist diagram
A -B
-
Real gain
Negative
imaginary gain
Cole cole plot
Summary of Contents for FRA5097
Page 1: ...NF Corporation FREQUENCY RESPONSE ANALYZER FRA5097 INSTRUCTION MANUAL ...
Page 2: ......
Page 3: ...DA00014283 002 FRA5097 Frequency Response Analyzer Instruction Manual ...
Page 4: ......
Page 35: ...2 2 Mounting and installation FRA5097 2 7 Fig 2 3 Removal of foot stands ...
Page 40: ......
Page 122: ......
Page 132: ......
Page 156: ......
Page 168: ......
Page 188: ......
Page 190: ......
Page 191: ......