3.2 Display at power “ON” and initial settings
FRA5097
3-15
Table 3-1 List of initialization values (continued)
Setting items
Initialization
values
At
shipment
from
factory
At data
destruction
At
initialization
executed
Notes/Remarks
OTHERS
TITLE SET
(Cleared)
○
○
−
Also cleared at
re-measurement
and at power
supply switch on
BUZZER ON
○
○
○
DATE SET
YEAR
1988
−
○
−
MONTH
1
−
○
−
DAY
1
−
○
−
The actual
shipment date is
set at shipment
from factory
TIME SET
HOUT
0
−
○
−
MINITE
0
−
○
−
The actual
shipment time is
set at shipment
from factory
3.2.3
Warm-up
It takes more than 30 minutes after the power supply switching-on for the internal temperature of
the
FRA5097
to reach stable.
Perform measurement right after calibration is made, which shall be made after sufficient time of
warm-up has been made. Note that the measurement accuracy specification is met under the
condition immediately after calibration.
Conduct re-calibration when environment temperature has been changed.