3.5 Examples of basic operations
FRA5097
3-30
3.5.4
Examples of connections
Typical examples of measurement connections are described in this section. In actual
measurement, be careful about the influence on the SUT of the common- and differential-mode
impedance of the oscillator output and analyzer input in the FRA5097.
Frequency characteristics measurement of amplifiers (1)
V
1
V
2
A
SUT
Fig. 3-14 Connection diagram for frequency characteristics measurement (1)
Connect V
1
with CH1 and V
2
with CH2, respectively.
The frequency transfer function of the SUT (system under test) can be obtained from the
following equation:
Set the measurement mode at (CH1 and CH2) and the analysis mode at (CH2/CH1) for the
FRA5097, to display the SUT frequency characteristics (gain and phase versus frequency) in the
LCD screen so that you can directly read them through display.
Frequency characteristics measurement of amplifiers (2)
V
1
V
2
SUT
A
Fig. 3-15 Connection diagram for frequency characteristics measurement (2)
The diagram above shows the connection for measurement of the transfer function of A (only for
A) in the SUT.
The transfer function of A can be obtained from the following equation:
Set the analysis mode at (CH2/CH1) for the FRA5097 to display the frequency characteristics
(gain and phase versus frequency) of A in the LCD screen so that you can
directly read them
through display.
A
=
V
2
V
1
・
・
A
=
V
2
V
1
・
・
Summary of Contents for FRA5097
Page 1: ...NF Corporation FREQUENCY RESPONSE ANALYZER FRA5097 INSTRUCTION MANUAL ...
Page 2: ......
Page 3: ...DA00014283 002 FRA5097 Frequency Response Analyzer Instruction Manual ...
Page 4: ......
Page 35: ...2 2 Mounting and installation FRA5097 2 7 Fig 2 3 Removal of foot stands ...
Page 40: ......
Page 122: ......
Page 132: ......
Page 156: ......
Page 168: ......
Page 188: ......
Page 190: ......
Page 191: ......