9.11 Miscellaneous specifications
FRA5097
9-14
9.11 Impedance display function
●
Display items
R: Impedance or admittance
A: Resistance or conductance
B: Reactance or susceptance
●
Additional graphs
Linear graphs (linear scale) and log graphs (log scale), 15 types
X axis
Y1 axis
Y2 axis
X axis
Y1 axis
Y2 axis
logF logR
θ
F logR
θ
logF
A B
F A B
logF logA logB
F
logA logB
logF
log(-A) logB
F
log(-A) logB
logF logA log(-B)
F
logA log(-B)
logF log(-A)
log(-B)
F
log(-A)
log(-B)
logF logR ---
F
logR ---
θ
logR
---
●
Scale unit
Gain (abstract number)/impedance (
Ω
, S)
●
Voltage and current input
CH1 fixed to voltage, CH2 fixed to current
●
Current shunt input conversion coefficient
0 to 1.0E+6 (resolution 5 digits or 0.01E-9), phase reverse feature
●
Open/short correction feature
Open correction
Short correction
Correction formula
OFF ON
Zx=Z-Zs
ON OFF Zx=Zp×Z÷(Zp-Z)
ON ON
Zx=Zp×(Z-Zs)÷(Zp-(Z-Zs))
Zx Correction calculation result
Z Measured value (CH1/CH2)
Zs Short correction data (CH1/CH2)
Zp Open correction data (CH1/CH2)
Calculation by the aforementioned formula applies regardless of the [ANALYSIS MODE]
(analysis mode) setting.
●
Maximum/minimum search feature
Search
item
Y1-axis maximum, Y1-axis minimum, Y2-axis maximum,
Y2-axis minimum
Search method
When the function key relevant to the search item is pressed when
data marker is being displayed, the vertical axis parameters are
automatically searched and the marker moves accordingly
Summary of Contents for FRA5097
Page 1: ...NF Corporation FREQUENCY RESPONSE ANALYZER FRA5097 INSTRUCTION MANUAL ...
Page 2: ......
Page 3: ...DA00014283 002 FRA5097 Frequency Response Analyzer Instruction Manual ...
Page 4: ......
Page 35: ...2 2 Mounting and installation FRA5097 2 7 Fig 2 3 Removal of foot stands ...
Page 40: ......
Page 122: ......
Page 132: ......
Page 156: ......
Page 168: ......
Page 188: ......
Page 190: ......
Page 191: ......