7.1 Error messages
FRA5097
7-4
No.
Messages
Descriptions
----------------------------------------
13 Can't load, file corrupted
The file is corrupted or is not correct, and
therefore it cannot be loaded.
14 Can't save during acquisition
Data is still in acquisition phase, and therefore
they cannot be saved.
15 Amplitude compression failure
Designated amplitude could not be set. Refer to
oscillator maximum output level and amplitude
variations, and review settings.
16 Storage area overflow
Memory capacity is not sufficient for storage and
overflow has occurred. Measurement data
cannot be stored anymore.
17 Can't set, range is too narrow
The sweep range which you are trying to set is
narrower than the frequency resolution capability.
You may not set this sweep range.
18 Can't set, line markers are invalid
Since the line markers in the y-axis direction are
invalid for the sweep range that you are setting,
you may not set the sweep range by using the
line markers.
19 Can't set, markers aren't displayed
Since markers are not displayed, you may not set
what you are trying to do.
20 Can't measure, analysis frequency is over limit
Analysis frequency is over 15 MHz which is the
upper limit, and therefore, measurement cannot
be done.
21 Can't set, line markers are invalid
Since the line markers in the y-axis direction are
invalid for the sweep range that you are setting,
you may not set the sweep range by using the
line markers.
23 Missing file name
The file name is missing.
24 Duplicate file name
Duplicated file names have been found.
Summary of Contents for FRA5097
Page 1: ...NF Corporation FREQUENCY RESPONSE ANALYZER FRA5097 INSTRUCTION MANUAL ...
Page 2: ......
Page 3: ...DA00014283 002 FRA5097 Frequency Response Analyzer Instruction Manual ...
Page 4: ......
Page 35: ...2 2 Mounting and installation FRA5097 2 7 Fig 2 3 Removal of foot stands ...
Page 40: ......
Page 122: ......
Page 132: ......
Page 156: ......
Page 168: ......
Page 188: ......
Page 190: ......
Page 191: ......