5.3 Open and short correction
FRA5097
5-8
5.3 Open and short correction
5.3.1 Connection for open and short correction data storage
Although open and short can be simultaneously corrected, these are used separately according to
the impedance to be measured.
Short correction: Low impedance (under about 10
Ω
)
Open correction: High impedance (above about 1 k
Ω
)
Connections for storing open and short correction data are indicated in Figs. 5-3 and 5-4
.
FRA5097
OSC
CH1 CH2
Measurement
terminal
open
Current detect
resistor
FRA5097
OSC
CH1 CH2
Measurement
terminal
Shorted
Current detect
resistor
Fig. 5-3 Open correction example Fig. 5-4 Short correction example
5.3.2 Data measurement and storage
Measure and store the open and short data as follows.
1) Set the menu [Measure] [BASIC FUNCTION] [OPEN CORRECTION] and [SHORT
CORRECTION] at off.
2) Measure the impedance terminal in open state. Use the menu [Memory] to select [STORAGE],
and so show the function keys indicated in Fig. 5-5. Press [OPEN STORAGE] to store the
open correction in memory.
MASS
STORAGE
PERMANENT
STORAGE
EQL
STORAGE
=
OPEN
STORAGE
SHORT
STORAGE
Fig. 5-5 Open and short correction function keys
3) Measure the terminal in shorted state and press the [SHORT STORAGE] key to store the short
correction data in memory.
Note: Be sure both correction functions are off when measuring the correction data. Measure the
DUT at the same frequency range, oscillator amplitude and other conditions as the correction
data.
Summary of Contents for FRA5097
Page 1: ...NF Corporation FREQUENCY RESPONSE ANALYZER FRA5097 INSTRUCTION MANUAL ...
Page 2: ......
Page 3: ...DA00014283 002 FRA5097 Frequency Response Analyzer Instruction Manual ...
Page 4: ......
Page 35: ...2 2 Mounting and installation FRA5097 2 7 Fig 2 3 Removal of foot stands ...
Page 40: ......
Page 122: ......
Page 132: ......
Page 156: ......
Page 168: ......
Page 188: ......
Page 190: ......
Page 191: ......