9.2 Analysis section input
FRA5097
9-5
Figure 9-1 indicates the isolation voltage specifications between the oscillating section (OSC)/analysis
section inputs (CH1 and CH2) and the enclosure.
OSC
CH1
CH2
、
*1
:
250 Vrms (AC), or
±
200V
(DC), alternatively
±
350
Vpeak (AC + DC)
250Vrms
250Vrms
250Vrms
250Vrms
Measurement
category I
250Vrms
*1
Measurement
category I
Fig. 9-1 From-enclosure isolation voltage specifications (when accompanying BNC
cable is used)
Figure 9-2 indicates the isolation voltage specifications when a cable other than the accompanying
cable is being used.
OSC
CH1
CH2
、
*2
:
30 Vrms (AC), or
±
60V
(DC), alternatively
±
42
Vpeak (AC + DC)
30Vrms
30Vrms
30Vrms
30Vrms
30Vrms
*2
Measurement
category I
Measurement
category I
Fig. 9-2 From-enclosure isolation voltage specifications (when a cable other than the
accompanying cable is used)
Figure 9-3 indicates the isolation voltage specifications between each of the oscillating section (OSC)
and analysis section inputs (CH1 and CH2) when the accompanying BNC cable is being used.
OSC
CH1
CH2
250Vrms
250Vrms
250Vrms
250Vrms
250Vrms
250Vrms
250Vrms
250Vrms
250Vrms
250Vrms
250Vrms
250Vrms
Fig. 9-3 Oscillation section input - analysis section input isolation voltage specifications
(when accompanying BNC cable is used)
Figure 9-4 indicates the isolation voltage specifications between each of the oscillating section (OSC)
and analysis section inputs (CH1 and CH2) when a cable other than the accompanying cable is being
used.
Summary of Contents for FRA5097
Page 1: ...NF Corporation FREQUENCY RESPONSE ANALYZER FRA5097 INSTRUCTION MANUAL ...
Page 2: ......
Page 3: ...DA00014283 002 FRA5097 Frequency Response Analyzer Instruction Manual ...
Page 4: ......
Page 35: ...2 2 Mounting and installation FRA5097 2 7 Fig 2 3 Removal of foot stands ...
Page 40: ......
Page 122: ......
Page 132: ......
Page 156: ......
Page 168: ......
Page 188: ......
Page 190: ......
Page 191: ......