8.5 Performance test
FRA5097
8-9
8.5.7 Analyzer IMRR
IMRR at 60 Hz is tested.
a)
Setting
FRA5097 oscillator
FRA5097 analyzer
Frequency
60 Hz
Integration
100 cycles
Output waveform
Sinewave
Measuring mode
CH1,CH2
Output voltage
Display mode
X axis : logF
AC
10 Vpeak
Y1 axis : dBR
DC bias
0 V
Y2 axis :
θ
b)
Connection
FRA5097
BNC-BNC
coaxial cable
BNC T-type
divider
BNC-alligator clip
coaxial cable
BNC shorting
plug
OSC
CH1 CH2
FRA5097
OSC
CH1 CH2
CH1 IMRR measurement CH2 IMRR measurement
c)
Procedure
Set the analyzer mode as follows. Conduct Single measurement with FRA5097.
Read the dBR measuring results from the LCD screen.
Analyzer mode for CH1 IMRR measurement: CH2/CH1
Analyzer mode for CH2 IMRR measurement: CH1/CH2
d)
Judgment
Connection Measured
value
Tolerance
CH1 (analyzer mode: CH2/CH1)
dBR= dB
More than 120 dB
CH2 (analyzer mode: CH1/CH2)
dBR= dB
More than 120 dB