3.3 Input and output terminals
FRA5097
3-16
3.3 Input and output terminals
Analyzer input terminals (CH1 and CH2)
CH1
CH2
2
ATT
1MΩ
cabinet
1
ATT
1MΩ
Each of the analyzer input terminals of the
FRA5097
is electrically insulated from the cabinet, the
oscillator and the other analyzer input terminal. The minimum breakdown voltage is 250 Vrms
(measurement category I) between each of the input and output terminals and the cabinet, between
CH1 and the oscillator, between CH2 and the oscillator and between CH1 and CH2, respectively
when the accompanying insulated coaxial cable is being used. Restriction of 30 Vrms applies
when a cable other than the accompanying cable is used. Note that accidents due to electric
shocks could occur, if voltages exceeding the minimum breakdown voltage are applied between
the above mentioned insulated parts, leading to dielectric breakdown. Refer to “3.4 Insulation
breakdown voltages of input and output voltages” by all means, when you make a measurement
where high voltages are applied between any two of the cabinet, CH1, CH2 and the oscillator.
WARNING
!
•
Do NOT connect to any measurement target that exceeds 250 Vrms of
measurement Category I.
Doing so may result in insulation breakdown, imposing electrical shock.
•
You could be suffered from electric shocks when you measure high voltage
circuit signals. Use accessory coaxial cables of the insulation type by all
means, so that you cannot directly touch metallic portions of the BNC
connectors at the analyzer input terminals.
The input terminals have the input impedance of 1 M
Ω
(parallel capacitance of 25 pF
±
5 pF) and
the maximum allowable input voltage of
±
350 V for AC+DC. Never apply any voltages
exceeding the maximum allowable voltage, since the inside of the instrument will be damaged by
application of voltages exceeding the minimum allowable voltage.
FRA5097
has the capability of measuring the amplitude and the phase of input signals up to 15
MHz. Use the same type and the same length of signal cables to be
connected to individual
channel inputs so that the phase can be measured with high accuracy at high frequencies.
The connection between the input connector and the internal circuits is cut off when the power
supply is off.
Summary of Contents for FRA5097
Page 1: ...NF Corporation FREQUENCY RESPONSE ANALYZER FRA5097 INSTRUCTION MANUAL ...
Page 2: ......
Page 3: ...DA00014283 002 FRA5097 Frequency Response Analyzer Instruction Manual ...
Page 4: ......
Page 35: ...2 2 Mounting and installation FRA5097 2 7 Fig 2 3 Removal of foot stands ...
Page 40: ......
Page 122: ......
Page 132: ......
Page 156: ......
Page 168: ......
Page 188: ......
Page 190: ......
Page 191: ......