2.3 Grounding and power supply connection
FRA5097
2-8
2.3
Grounding and power supply connection
Ground the instrument by all means.
!
WARNING
Line filters are used for this instrument, and therefore, you will have a risk of
an electric shock unless the instrument is properly grounded.
To avoid risk of electric shock, be sure to connect securely to ground through
less than 100
Ω
.
This instrument is so designed that the instrument will be grounded by connecting its three-pole
power supply plug with a three-pole electric power source outlet with a proper grounding
connection.