9.2 Analysis section input
FRA5097
9-4
9.2 Analysis section input
●
N
umber of output channels
2 channels
●
Connector
Insulated BNC connector
●
Input impedance
1 M
Ω
±
2%, parallel 25 pF
±
5 pF
●
IMRR (isolation mode removal ratio)
120 dB or above (DC up to 60 Hz)
This value applies when the signal source impedance is smaller than 1
Ω
.
●
Isolation
Withstand voltage
250 Vrms continuous (between signal/ground - enclosure)
250 Vrms continuous (between signal/ground - analysis section input,
between the analysis section channels)
This value applies when the accessory BNC cable is being used.
30 Vrms applies when a cable other than the accessory cable is used.
Against-enclosure capacity No more than 200 pF
●
Measurement category
I, maximum transient voltage 1,500 Vrms
●
Frequency range
0.1 mHz to 15 MHz
●
Maximum input voltage
250 Vrms (AC) or
±
200 V (DC) alternatively
±
350 Vpeak (AC + DC)
This value applies when accessory BNC cable is used
When a cable other than the accessory cable is used, 30 Vrms
(AC), or
±
60 V (DC)
alternatively
±
42 Vpeak (AC + DC)
●
Maximum measurement
250Vrms
voltage
This value applies when the accessory BNC cable is being used.
30 Vrms applies in case a cable other than the accessory cable is used.
●
Over detection
Setting range
0 to 250 Vrms
Setting resolution
3 digits
Processes
Over lamp lights
Warning
sound
(can be turned ON/OFF)
Sweep abort (can be enabled/disabled)
Oscillator OFF (can be enabled/disabled)
●
Harmonic measurement
2nd to 10th (highest frequency of the harmonics is 15 MHz)
●
Harmonic and noise rejection ratio
Normal mode DC
60 dB or above
Broadband white noise
50 dB or above (noise bandwidth 500 kHz, integration of 1,000 cycles)
Harmonic (10th and below) 60 dB or above (analysis frequency 100 kHz or below)
40 dB or above (analysis frequency 100 kHz or below)
●
Dynamic range
140 dB typ (10 Hz to 1 MHz)
80 dB typ. (above 1MHz, no more than 15 MHz)
(This value applies when the bigger of the channel inputs is at least 10
Vpeak, with an integration of 4,000 cycles)
●
Input weight
0 to 1.0E+6 (resolutions 5 digits or 0.01E-9)
Summary of Contents for FRA5097
Page 1: ...NF Corporation FREQUENCY RESPONSE ANALYZER FRA5097 INSTRUCTION MANUAL ...
Page 2: ......
Page 3: ...DA00014283 002 FRA5097 Frequency Response Analyzer Instruction Manual ...
Page 4: ......
Page 35: ...2 2 Mounting and installation FRA5097 2 7 Fig 2 3 Removal of foot stands ...
Page 40: ......
Page 122: ......
Page 132: ......
Page 156: ......
Page 168: ......
Page 188: ......
Page 190: ......
Page 191: ......