5.3 Open and short correction
FRA5097
5-9
5.3.3 Open and short correction data memory
Battery backup is not used for open and short correction data memory. These data are lost when
power is cut and resupplied. Be sure to again enter the data after supplying power.
The capacity sum of the master and equalizer memories corresponds to 20,000 measurements. If
large amounts of data are stored in the master memory, the amount available for the correction
data is reduced and in some cases may be inadequate. In this event, erase the master memory
before storing the correction data.
5.3.4 Open and short correction function setting
The open and short correction function must be set to on in order to conduct measurements using
the open and short correction data. Use the menu [MEASURE] [BASIC FUNCTION] [OPEN
CORRECTION] and [SHORT CORRECTION] to select and deselect (on and off).
The
items can be selected independently. Table 5-4 shows the correction formulas for the item
on/off state combinations.
Table 5-4 Open and short correction formulas
Open correction
Short correction
Correction formula
OFF ON
Zx=Z-Zs
ON OFF
Zx=Zp
×
Z
÷
(Zp-Z)
ON ON
Zx=Zp
×
(Z-Zs)
÷
(Zp-(Z-Zs))
Zx: Correction computation result
Z : Measurement value
Zs: Short correction data (CH1/CH2)
Zp: Open correction function (CH1/CH2)
These formulas apply regardless of the [ANALYSIS MODE].
Summary of Contents for FRA5097
Page 1: ...NF Corporation FREQUENCY RESPONSE ANALYZER FRA5097 INSTRUCTION MANUAL ...
Page 2: ......
Page 3: ...DA00014283 002 FRA5097 Frequency Response Analyzer Instruction Manual ...
Page 4: ......
Page 35: ...2 2 Mounting and installation FRA5097 2 7 Fig 2 3 Removal of foot stands ...
Page 40: ......
Page 122: ......
Page 132: ......
Page 156: ......
Page 168: ......
Page 188: ......
Page 190: ......
Page 191: ......