4.2 Delaying measurement start
FRA5097
4-7
4.2 Delaying measurement start
When the oscillator frequency is changed in the frequency sweep mode, the measurement data
may, in general, have errors due to transient response if the SUT (system under test) involves any
response time delay components. The measurement start time is set with some delay time period
for the FRA5097 frequency response characteristics measurement, so that the transient error can
be suppressed at the minimum level. You can set the delay time period at an arbitrary value
according to the time constant of the SUT.
Among others, note that sufficient period of time delay needs to be secured for measuring
accurately the SUTs that involve high order elements (e.g., filters with steep frequency
characteristics, high Q resonance circuits with crystal oscillators, etc.), since they have long period
of time delay.
Use the menu [Measure][BASIC FUNCTION][DELAY TIME] to set the period of time delay.
The period of time delay can be set with respect to either the number of cycles of analysis
frequency (unit: cycle) or the time period (unit: second).
OSC frequency = f1
f2
Measurement time range
for the time delay
= 0.
Measurement time
range for the proper
amount of time delay.
time
OSC output
waveform
SUT response
waveform
Fig. 4-4 Response waveform requiring measurement start time delay
Summary of Contents for FRA5097
Page 1: ...NF Corporation FREQUENCY RESPONSE ANALYZER FRA5097 INSTRUCTION MANUAL ...
Page 2: ......
Page 3: ...DA00014283 002 FRA5097 Frequency Response Analyzer Instruction Manual ...
Page 4: ......
Page 35: ...2 2 Mounting and installation FRA5097 2 7 Fig 2 3 Removal of foot stands ...
Page 40: ......
Page 122: ......
Page 132: ......
Page 156: ......
Page 168: ......
Page 188: ......
Page 190: ......
Page 191: ......