9.1 Oscillation section
FRA5097
9-3
●
Frequency sweep
Sweep range
0.1 mHz to 15 MHz
Sweep density
Log sweep
:
3 to 20,000 steps/sweep or 1 to 20,000 steps/decade
(Note 3 steps sweep minimum, 20,000 steps/sweep maximum)
Linear sweep
:
3 to 20,000 steps/sweep or 0.1 to 15 MHz
(Note 3 steps sweep minimum, 200,000 steps/sweep maximum)
●
Isolation
Withstand voltage
250 Vrms continuous (between signal/ground - enclosure)
250 Vrms continuous (between signal/ground - analysis section
input)
This value applies when the accompanying BNC cable is
being used.
30 Vrms continuous applies when a cable other than the
accessory cable is used.
Against-enclosure capacity No more than 250 pF
●
Measurement category
I, maximum transient voltage 1,500 Vrms