8.5 Performance test
FRA5097
8-5
8.5.3 Oscillator output frequency accuracy
The test of oscillator frequency accuracy is described in this paragraph.
a)
Setting
FRA5097 Frequency counter
Output waveform sinusoidal wave
Gate time
10 sec.
Output voltage
AC:
1 Vpeak
DC bias:
0 V
b)
Connection
FRA5097
Frequency counter
BNC-BNC
Coaxial cable
OSC
CH1 CH2
c)
Procedure
Set the oscillator output frequency at the following and read the indication of the frequency
counter:
Oscillator output frequency: 100 kHz
d)
Result and judgment
Oscillator frequency
Frequency counter
Permissible range
100 kHz
kHz
99.9990 - 100.0010 kHz