9.1 Oscillation section
FRA5097
9-2
9.1 Oscillation section
●
Number of output channels
1 channel
●
Connector
Insulated BNC connector
●
Output waveform
Sine wave/rectangular wave/triangular wave
●
Frequency
Range
0.1 mHz to 15 MHz (In the case of rectangular or triangular wave,
be sure to use at 100 kHz or below. The waveform quality
degrades above 100 kHz.)
Setting resolution
0.1mHz
(Resolution is 1 mHz at 10 MHz or higher.)
Accuracy
±10ppm
●
AC amplitude
Range
0V to 10 Vpeak (under no load)
Setting resolution
3 digits or 0.01 mVpeak, whichever is greater
Accuracy (sine wave)
Within
±
0.3 dB (100 kHz or below)
Within
±
1 dB (1 MHz or below)
Within
±
3 dB (15 MHz or below)
(The above assumes a calibration was performed shortly prior, and
that the setting is between 100 mV to 10 Vpeak.)
Distortion (sine wave)
No more than 0.2% (when 100 kHz or below, BW 500 kHz, and
10 Vpeak output)
●
DC bias
Range
-10V to 10V (under no load)
Resolution
10mV
Accuracy
±
(1% of DC bias s 2% of AC amplitude s 30mV)
(This value is applicable if a calibration was done shortly prior/)
●
Output impedance
50
Ω
±2%(when 1 kHz) Unbalanced (BNC connector plug)
●
Maximum output (AC+DC)
Voltage
±10V(no load)
Current
±100mA
●
Output control
QUICK (instantly changes to the set voltage or to 0 V)
SLOW (gradually increases/decreases until reaching the set voltage or 0 V)
Phase control (The oscillation start/stop phase can be set by one degree increments.)
AC and DC can be simultaneously turned ON/OFF, while AC can be independently turned OFF.
Summary of Contents for FRA5097
Page 1: ...NF Corporation FREQUENCY RESPONSE ANALYZER FRA5097 INSTRUCTION MANUAL ...
Page 2: ......
Page 3: ...DA00014283 002 FRA5097 Frequency Response Analyzer Instruction Manual ...
Page 4: ......
Page 35: ...2 2 Mounting and installation FRA5097 2 7 Fig 2 3 Removal of foot stands ...
Page 40: ......
Page 122: ......
Page 132: ......
Page 156: ......
Page 168: ......
Page 188: ......
Page 190: ......
Page 191: ......