8.5 Performance test
FRA5097
8-10
8.5.8 Analyzer dynamic range
The dynamic range is tested.
a)
Setting
FRA5097 oscillator
FRA5097 analyzer
Sweep upper limit frequency
15 MHz
Integration
2,000 cycles
Sweep lower limit frequency
10 Hz
Measuring mode
CH1,CH2
Output waveform
Sinewave
Display mode
X axis :
logF
Output voltage
Y1 axis: dBR
AC
10 Vpeak
DC bias
0 V
b)
Connection
FRA5097
BNC-BNC
coaxial cable
BNC
shorting plug
OSC
CH1 CH2
FRA5097
OSC
CH1 CH2
CH1 dynamic range measurement
CH2 dynamic range measurement
c)
Procedure
Set the analyzer mode as follows and sweep with the FRA5097.
Read the minimum dBR measuring results from the LCD screen.
Analyzer mode for measuring CH1 dynamic range: CH2/CH1
Analyzer mode for measuring CH2 dynamic range: CH1/CH2
d)
Judgment
Connection
Minimum measured value
Tolerance
CH1 (10 Hz to 1 MHz)
dBR= dB
140 dB typ
CH2 (10 Hz to 1 MHz)
dBR= dB
140 dB typ
CH1 (1 MHz to 15 MHz)
dBR= dB
80 dB typ
CH2 (1 MHz to 15 MHz)
dBR= dB
80 dB typ