8.5 Performance test
FRA5097
8-8
8.5.6 Oscillator output DC bias accuracy
The oscillator output DC bias accuracy is tested.
a)
Setting
FRA5097
Multi-meter
Output waveform
Sinusoidal wave
Measuring mode DC voltage
Output voltage
AC
0 Vpeak
b)
Connection
FRA5097
Multi-meter
BNC-BNC
Coaxial cable
OSC
CH1 CH2
c)
Procedure
Set the oscillator DC bias as follows and read the multi-meter indication.
Oscillator DC bias: -10, 0, +10 V
d)
Judgment
Oscillator DC bias
Multi-meter
Tolerance
-10V
V
-10.130 to –9.870V
0V
V
-30.0 to +30.0mV
+10V
V
+9.870 to +10.130V