6.3 File format
FRA5097
6-5
Table 6-2 b) Measurement data file format - Set parameters
Offset Type
Descriptions
(content)
40(0x28)
long
(Not yet used)
44(0x2c)
short
Type of data (data type) 0: RAW data, 1: OPERATED data
46(0x2e)
short
Number of data (Num)
48(0x30) short
Data status
0: not stored
1: stored in mass memory
2: stored in permanent memory
50(0x32)
short
Year when data are generated:
1970 - 2069
52(0x34)
short
Month when data are generated:
1 -12
54(0x36)
short
Day when data are generated:
1 - 31
56(0x38)
short
Hour when data are generated:
0 - 23
58(0x3a)
short
Minute when data are generated:
0 - 59
60(0x3c)
char[64] Title of data 64 characters maximum
124(0x7c)
double
Oscillator (OSC) amplitude 0.0 - 10.0
132(0x84)
double
Oscillator (OSC) bias -10.0 - +10.0
140(0x8c) short
Oscillator (OSC) waveform 0:sinusoidal wave,
1:rectangular wave, 2:triangular wave
142(0x8e)
short
(not yet used)
144(0x90)
double
Maximum sweep frequency
0.1E-3 - 15E+6
152(0x98)
double
Minimum sweep frequency
0.1E-3 - 15E+6
160(0xa0) short
Type of sweep resolution (Sweep Type)
0:Log steps/sweep 2:Lin
steps/sweep
1:Log steps/decade
3:Lin Hz
162(0xa2)
short
(not yet used)
SweepType=0,1,2
SweepType=3
164(0xa4)
short
Sweep resolution (steps)
166(0xa6)
short[3]
(not yet used)
double
Sweep resolution (Hz)
172(0xac) short Type
of
integration (IntegType) 0:cycle, 1:sec
174(0xae)
short
(not yet used)
IntegType=0
IntegType=1
176(0xb0) short
Number of integrations
(cycle)
178(0xb2)
short[3]
(not yet used)
double Integration
time
(sec)
184(0xb8)
short
Type of delay (Delay Type) 0:cycle, 1:sec
186(0xba)
short
(not yet used)
DelayType=0
DelayType=1
188(0xbc)
short
Delay period (cycle)
190(0xbe)
short[3]
(not yet used)
double
Delay time (sec)
196(0xc4)
short
Order of harmonics 1 - 10
198(0xc6) short Measurement
mode
0:CH1&CH2, 1:CH1&OSC, 2:OSC&CH2
200(0xc8) long Automatic integration function
0:OFF, 1:ON
204(0xcc) long High-density slow sweep function 0:OFF, 1:ON
208(0xd0) long Amplitude compression function
0:OFF, 1:ON
The file format for the data depends on the types of data (i.e., RAW or OPERATED data). The
type of data is specified by the DataType of the offset 44(0x2c) in the set parameter. The first
one (point of) data for either of the RAW or OPERATED data (RAW[0] or OPRD[0]) is invalid.
The number of data is stored in Num in the offset 46(0x2e) of the set parameter.
Summary of Contents for FRA5097
Page 1: ...NF Corporation FREQUENCY RESPONSE ANALYZER FRA5097 INSTRUCTION MANUAL ...
Page 2: ......
Page 3: ...DA00014283 002 FRA5097 Frequency Response Analyzer Instruction Manual ...
Page 4: ......
Page 35: ...2 2 Mounting and installation FRA5097 2 7 Fig 2 3 Removal of foot stands ...
Page 40: ......
Page 122: ......
Page 132: ......
Page 156: ......
Page 168: ......
Page 188: ......
Page 190: ......
Page 191: ......