3.5 Examples of basic operations
FRA5097
3-28
b)
Example of REPEAT measurement
Repeat the SINGLE measurement (measurement at a fixed frequency). Press the
REPEAT
key to turn on the
REPEAT
lamp, and then press the
MEAS
key to obtain the measurement
data.
The measurement data are displayed in the center of the LCD screen as is the case for the
SINGLE measurement. The measurement data display is updated at every time when one
time of measurement is completed.
c)
Example of Sweep measurement
The Sweep measurement is a measurement mode where measurement is made through
frequency sweep. An example of measurement is described below with the following
measurement conditions:
・
Measurement frequency
: 1.0 kHz - 2.2 MHz
・
Oscillator
:
Amplitude=7
Vpeak
・
Number of times of integration : 10 cycle
The following parameters are unchanged from those at completion of initialization:
・
Sweep resolution
: LOG 100 steps/sweep
・
Oscillation waveform
: sinusoidal
・
Delay time
: 0 second
・
Measurement mode
: CH1 and CH2
・
Analysis mode
: CH1/CH2
・
Display mode
: X-axis : logF
Y1-axis
:
dBR
Y2-axis
:
θ
1)
Connection with the SUT (system under test)
Connect the FRA5097 with the SUT (system under test) according to
“Fig. 3-12
Connection with SUT”
for the measurement mode of (CH1 and CH2).
Summary of Contents for FRA5097
Page 1: ...NF Corporation FREQUENCY RESPONSE ANALYZER FRA5097 INSTRUCTION MANUAL ...
Page 2: ......
Page 3: ...DA00014283 002 FRA5097 Frequency Response Analyzer Instruction Manual ...
Page 4: ......
Page 35: ...2 2 Mounting and installation FRA5097 2 7 Fig 2 3 Removal of foot stands ...
Page 40: ......
Page 122: ......
Page 132: ......
Page 156: ......
Page 168: ......
Page 188: ......
Page 190: ......
Page 191: ......