8.5 Performance test
FRA5097
8-6
8.5.4 Oscillator output amplitude accuracy
The oscillator output amplitude and frequency response are tested.
a)
Setting
FRA5097
Multi-meter
Output waveform sinusoidal wave
Measurement mode
AC voltage
Output voltage
(True
RMS)
AC
10 Vpeak
DC bias
0V
b)
Connection
FRA5097
Multi-meter
BNC-BNC
Coaxial cable
OSC
CH1 CH2
c)
Procedure
Set the oscillator frequency as follows and read the multi-meter indications. Use a
wideband multi-meter for frequencies above 100 kHz.
Oscillator frequency: 1 kHz, 100 kHz, 1 MHz, 15 MHz
Derive the amplitude from the multi-meter indication by the following formula.
[Amplitude accuracy (dB) = 20
×
log10[multi-meter indication (Vrms)] - 16.9897
d)
Judgment
Oscillator frequency
Multi-meter
Amplitude accuracy
Tolerance range
1kHz
Vrms
dB -0.3
–
+0.30dB
100kHz
Vrms
dB -0.3
–
+0.30dB
1MHz
Vrms
dB -1.00
–
+1.00dB
15MHz
Vrms
dB -3.00
–
+3.00dB