8.5 Performance test
FRA5097
8-11
8.5.9 Analyzer measuring error frequency response
Frequency dependence of the CH1/CH2 measurement value at 100 mVpeak output is tested.
a)
Setting
FRA5097 oscillator
FRA5097 analyzer
Sweep upper limit frequency
15 MHz
Integration
50 cycles
Sweep lower limit frequency
10 Hz
Measuring mode
CH1, CH2
Output waveform
Sinewave
Analyzing mode
CH1/CH2
Output voltage
Display mode
X axis : logF
AC
100 mVpeak
Y1
axis
: dB
R
DC bias
0 V
X2 axis :
θ
b)
Connection
FRA5097
BNC-BNC
coaxial cables
BNC
T-type divider
OSC
CH1 CH2
c)
Procedure
Use FRA5097 for sweep measurement.
Read the absolute maximum measured results for dBR and
θ
for frequency ranges up to 20
kHz, 500 kHz, 2.2 MHz and 15 MHz from the LCD screen.
d)
Judgment
Frequency range
Measured value
Tolerance
dB -0.05to
+0.05
dB
10 Hz to 20 kHz
deg
-0.3 to +0.3 deg
dB
-0.1 to +0.1 dB
20 kHz to 500 kHz
deg
-0.5 to +0.5 deg
dB
-1.0 to +1.0 dB
500 kHz to 2.2 MHz
deg
-2.0 to +2.0 deg
dB
-2.0 to +2.0 dB
2.2 MHz to 15 MHz
deg
-5.0 to +5.0 deg
Summary of Contents for FRA5097
Page 1: ...NF Corporation FREQUENCY RESPONSE ANALYZER FRA5097 INSTRUCTION MANUAL ...
Page 2: ......
Page 3: ...DA00014283 002 FRA5097 Frequency Response Analyzer Instruction Manual ...
Page 4: ......
Page 35: ...2 2 Mounting and installation FRA5097 2 7 Fig 2 3 Removal of foot stands ...
Page 40: ......
Page 122: ......
Page 132: ......
Page 156: ......
Page 168: ......
Page 188: ......
Page 190: ......
Page 191: ......