Ventilator Control and Drive
Theory of Operation
1 - 18
T23
-10V
Negative Voltage -10V
-10V
T24
A12V
Analogue 12V
12V
T25
VCC
Power 5V
5V
T26
5VAUX
5V Auxiliary
5V
T27
10V
An10V
+10V
T28
AD-Inhale
Inhale-STATE feedback
No activity may indicate
open connection from driver
to valve (ie. bad cable)
T29
AGND
Analogue Ground
T30
GND
Ground
T31
PAW
Pressure inside air way
0.45-0.55 (open to
atmosphere)
T32
Manual
Manual/Auto Valve feedback
Test Point Definition
DESIGNATOR
NAME
FUNCTION
RANGE