Preoperative Checklist
Periodic Maintenance
6 - 44
19. Test the Tidal Volume in the PCV ventilation mode:
a.
Set the O
2
flow to 2 L/min and set the N
2
O and AIR flow rates to minimum flow.
b.
Set the ventilator controls to:
c.
Select
PCV
to begin ventilation.
d.
Verify that the PEAK Pressure settles within ±4 of the set value within 3 breaths.
e.
Re-activate
CMV
ventilation mode.
20. Low FiO
2
Alarm Test
a.
Set the low FiO
2
Alarm limit to 50% O
2
.
b.
Set the AIR flow control valve to 5 L/min.
c.
Set the FiO
2
flow controller to minimum flow.
d.
Verify the following Low FiO
2
alarm signals activate, within three ventilation cycles:
•
Low FiO
2
message appears on the screen.
• An alarm tone sounds.
e.
Set the Low FiO
2
alarm limit to 18%.
• Verify the alarm signals stop activating.
21. High FiO
2
Alarm Test
a.
Set the high FiO
2
Alarm limit to 49% O
2
.
b.
Set the FiO
2
flow control valve to 5 L/min.
c.
Set the AIR flow controller to minimum.
d.
Verify the following High FiO
2
alarm signals activate:
•
High FiO
2
message appears on the screen.
• An alarm tone sounds.
e.
Set the high FiO
2
alarm limit to the max setting.
• Verify the alarm signals stop activating.
ATTRIBUTE SETTING
Patient Type
Adult
Ventilation Mode
PCV
Target Pressure -
P
TARGET
20
Breathng Rate -
freq
8
I:E Ratio -
I:E
1:2
PEEP -
PEEP
Off
Inspiratory Slope -
Tslope
0.5
Summary of Contents for Datascope AS 3000
Page 2: ...AS Datascope Anesthesia Delivery System Service Manual...
Page 9: ...Table of Contents vi 0070 10 0683 AS3000 Service Manual This page intentionally left blank...
Page 13: ...Introduction Notes x 0070 10 0683 AS3000 Service Manual This page intentionally left blank...
Page 179: ...0070 10 0683 Rev K August 8 2014...