Functional Tests
Periodic Maintenance
6 - 22
6.8.4.6
Oxygen Sensor Calibration
NOTE:
Oxygen Sensor Calibration can be performed in all
ventilation modes.
1.
Press the
MENU
key and then use the
Navigator
™
Knob
to scroll to the
Calibrate
menu tab (see FIGURE 6-22). Select the
Start Calibration
button.
FIGURE 6-22 Calibrate
Menu Tab
2.
After the
Start Calibration
button has been selected, the screen shown in
FIGURE 6-23 will be displayed, instructing the user to remove the oxygen sensor from
the Breathing System and expose it to room air for at least 3 minutes before proceeding.
After at least 3 minutes have passed, select the
Next
button to initiate the calibration
process. The progress bar shown in FIGURE 6-24 will be displayed.
NOTE:
If the system is going to be used during the calibration,
insert the O
2
cell plug into the port from which the oxygen
sensor was removed using a push and turn motion.