Installation Checkout Procedure
Installation Guide
2 - 8
2.3
Installation Checkout Procedure
Complete each step to verify the functionality of the
prior to clinical use.
Also, perform this checkout after installation, reinstallation, servicing or after any periodic
maintenance activity. This checklist does not replace periodic maintenance actions that must
be performed to maintain peak performance.
1. Verify that all components are present, and inspect for physical damage.
a.
Ensure that the Operating Instructions is present.
b.
Verify unit is free from cosmetic defects. Plastics, labeling and display window are
free from inclusions, pitting, bulges, sink marks, nicks, scratches, gouges dents,
discolorations, etc. Verify all major hardware is fastened properly.
c.
Verify that there are Tank Washers for each of the external cylinder yokes.
d.
Verify that each vaporizer's locking spring is intact on the manifold by looking
through the opening and checking for the visible wire spring.
e.
Verify that each connector of the vaporizer mount has an O-ring.
f.
Verify that breathing circuit consumables and CO
2
absorbent are present.
g.
Inspect the O
2
, N
2
O, AIR, VAC, and EVAC supply hoses for damages.
h.
Turn the flow control knobs for O
2
, AIR and N
2
O and verify that the floats rise and
spin as the flow is increased and fall freely as the flow is decreased.
i.
inspect the AC line cord for fraying and damages.
2. Verify that the laminated Preoperative Checkout card is attached.
3. Verify that the tank wrench is attached.
4. Verify that the waste scavenger flow rate is set and that the hose is
intact, connected and dry.
a.
Attach the waste gas hose from the waste gas port on the Breathing System to one of
the waste gas ports on the AGSS.
b.
Attach the EVAC hose from the waste gas assembly DISS (Diameter Index Safety
System) fitting to a source of vacuum.
c.
Adjust the position of the float to be between the Min and Max lines by turning its
flow adjustment knob (counterclockwise increases flow, clockwise decreases flow).
Summary of Contents for Datascope AS 3000
Page 2: ...AS Datascope Anesthesia Delivery System Service Manual...
Page 9: ...Table of Contents vi 0070 10 0683 AS3000 Service Manual This page intentionally left blank...
Page 13: ...Introduction Notes x 0070 10 0683 AS3000 Service Manual This page intentionally left blank...
Page 179: ...0070 10 0683 Rev K August 8 2014...