1 - 27
Theory of Operation
Ventilator Control and Drive
Power Board
Battery input, J4
PIN NAME FUNCTION
1. BAT
Battery
Input
+
2. 5V
Power
Ground
3. BATTEST
Battery
test
input
DC24V Power input, J3
PIN NAME FUNCTION
4. GND Power
Ground
5. GND
Power
Ground
6. 24V
Power
input
7. 24V
Power
input
DC Power output, J2
PIN NAME FUNCTION
1. GND Power
Ground
2.
+5V
Controller Logic Power
3. P12V Power
12V
4.
PWBUS
Power Bus 24V
5. GND Analog
Ground
6. A12V Analog
12V
7. 5V/4.8V
Power
5V/4.8V
Heater Power, J1
PIN NAME FUNCTION
6. GND Power
Ground
7.
5V
Controller Power, Output Enable
8.
24V
Heater Power Supply
9.
PBUS
Power Bus, No Connection
Power Control, J2
PIN NAME FUNCTION
1. GND Ground
2.
/PWON
Power On Enable, Low active
3.
PWEN
Power On Enable, High active
4. 5VAUX 5V
Auxiliary
5.
DC_M
AC/DC Monitor signal
6. PWBUS Power
bus
7.
BAT_M
Battery Monitor signal
Summary of Contents for Datascope AS 3000
Page 2: ...AS Datascope Anesthesia Delivery System Service Manual...
Page 9: ...Table of Contents vi 0070 10 0683 AS3000 Service Manual This page intentionally left blank...
Page 13: ...Introduction Notes x 0070 10 0683 AS3000 Service Manual This page intentionally left blank...
Page 179: ...0070 10 0683 Rev K August 8 2014...