5 - 3
Calibration
Calibration Warnings, Precautions, and Notes
5.2.3
Notes
NOTE:
Only bacterial filters with a low flow resistance must be
connected to the patient module and/or the patient
connection.
NOTE:
Use surgical gloves whenever touching or disassembling
valves or other internal components of the Breathing
System.
NOTE:
Ensure that the gas supply of the machine always complies
with the technical specification.
NOTE:
The APL Valve and PAW gauge numerics are for reference
only. Calibrated patient airway pressure is available on the
ventilator screen.
NOTE:
If the machine should show faults during the initial
calibration or testing, the machine should not be operated
until the fault has been repaired by a qualified service
technician.
NOTE:
After servicing, functional, sensor and system tests must be
carried out before clinical use.
NOTE:
To accommodate additional monitors and other equipment
the AS3000 offers up to two vertical mounting tracks. Use
of unauthorized mounting accessories is not recommended.
NOTE:
Always secure any equipment placed on the AS3000’s top
shelf.