5 - 5
Calibration
Calibration Procedures
NOTE:
If the system is going to be used during the calibration,
insert the O
2
cell plug into the port from which the oxygen
sensor was removed using a push and turn motion.
a.
Select
21%
to enter the 21% oxygen concentration calibration screen.
b.
Remove the oxygen sensor from the Breathing System and expose it to room air for
at least 3 minutes.
c.
Flush the O
2
sensor with air from the auxiliary output for 5-10 seconds to ensure that
no O
2
bubbles are trapped in the sensor.
d.
Select
Next
to start 21% oxygen concentration calibration.
NOTE:
The O
2
sensor voltage is displayed during the calibration.
This is the amplified O
2
cell voltage at the A/D converter for
the oxygen sensor. The O
2
sensor voltage is not displayed
for UI versions 2.24 and lower.
e.
When calibration is successfully completed, install the oxygen sensor into the
Breathing System.
NOTE:
The oxygen sensor must be installed in the Breathing
System for 10 minutes prior to 100% oxygen concentration
calibration to adjust to the temperature of the system.
f.
Press the O
2
Flush button for 5-10 seconds to clear out any non-O
2
gases from the
system.
g.
Expose the oxygen sensor to 100% pure oxygen (5 L/min) for at least 3 minutes.
h.
Select
100%
to enter the 100% oxygen concentration calibration screen.
i.
Select
Next
to start 100% oxygen concentration calibration.
Summary of Contents for Datascope AS 3000
Page 2: ...AS Datascope Anesthesia Delivery System Service Manual...
Page 9: ...Table of Contents vi 0070 10 0683 AS3000 Service Manual This page intentionally left blank...
Page 13: ...Introduction Notes x 0070 10 0683 AS3000 Service Manual This page intentionally left blank...
Page 179: ...0070 10 0683 Rev K August 8 2014...