5 - 9
Calibration
Calibration Procedures
5.5.4
Flow Valve Calibration
1. Preparing the unit
FIGURE 5-5
Connecting the breathing circuit
a.
Reinstall the Inspiratory and Expiratory valve disks.
b.
Connect a breathing circuit to the Breathing System (see FIGURE 5-5).
c.
Connect a 3L bag to the Y-fitting.
NOTE:
A 2.3L bag can be used in place of a 3L bag.
d.
Select
Continue.
2. Calibration
a.
Press the O
2
Flush Valve to fully fill bellows.
b.
Select
Continue
to start calibration.
c.
Follow the on-screen instructions.
NOTE:
The flow Valve calibration duration is approximately 5
minutes.