1 - 15
Theory of Operation
Ventilator Control and Drive
O
2
Sensor Input, S4
PIN NAME
FUNCTION
1. O
2
-
O
2
Sensor Input -
2. O
2
-
O
2
Sensor Input -
3. O
2
+
O
2
Sensor Input +
4. O
2
+
O
2
Sensor Input +
DC Power Input, J7
PIN NAME
FUNCTION
1. GND
Power
Ground
2.
+5V
Controller Logic Power
3. P12V
Power
12V
4.
PWBUS
Power Bus 24V
5. GND
Analog
Ground
6. A12V
Analog
12V
7. 5V/4.8V
Power
5V/4.8V
Power Control, J20
PIN NAME
FUNCTION
1. GND
Ground
2.
/PWON
Power On Enable, Low active
3.
PWEN
Power On Enable, High active
4. 5VAUX
5V
Auxiliary
5.
DC_M
AC/DC Monitor signal
6. PWBUS
Power
bus
7.
BAT_M
Battery Monitor signal
8.
APE
Auxiliary Power Enable
9.
CHAR_SIGNAL
Charge Status, No connection
10.
/MUTE
Mute, No connection
Signals and Power to Keyboard, J5
PIN NAME
FUNCTION
1.
ENPW
Power Enable, High active
2.
/PWON /
Power on Enable, Low active
3. 5VAUX
5V
Auxiliary
4.
APE
Auxiliary Power Enable
5. GND
Ground
6. TTLRX
TTL
Receive
7. 232ARX RS232A
Receive
8. 232BRX RS232B
Receive
9. GND
Ground
Summary of Contents for Datascope AS 3000