2 - 3
Installation Guide
Assembly
2.2.2.1
Attaching the Mounting Arms and User Interface
1.
Remove the mounting arms from small box and install them onto the
AS3000
using a 3
mm allen wrench.
2.
Remove the keys from the parts box, and unlock the
AS3000
drawers and put parts
box into top drawer.
3.
Remove the AGSS and Breathing System mount from the foam packaging, and install
them onto the
AS3000
.
4.
Remove the User Interface and slide it onto the User Interface mounting arm.
5.
Apply Loctite 243 to the User Interface cable jackscrews and jackposts. Then, connect
the User Interface cable to the User Interface. Ensure the jackscrews are securely
tightened.
2.2.2.2
Breathing System Connections
1.
Remove the foam and bag from around the Breathing System.
2.
Ensure that the serial number tag attached to the bag arm matches the serial number
found on the package carton.
3.
Install the Breathing System onto the arm on the left side of the
AS3000
by attaching it
to the installed bracket.
4.
Attach the Drive Gas Hose to the bottom port of the Breathing System
5.
Attach the Heater cable into the port located on the bottom of the Breathing System
6.
Install the O
2
Cell to the O
2
Interface
7.
Attach the O
2
cell to the O
2
cable and insert the cell into its designated port on the
Breathing System.
8.
Attach the Pneumatic Hose Assembly to the designated port on the rear of the Breathing
System.
9.
Attach the AGSS Transfer Hose to its designated port on the bottom of the Breathing
System.
2.2.2.3
Anesthesia System Connections
1.
Attach the Fresh Gas Hose of the Breathing System to the CGO port on the system.
Summary of Contents for Datascope AS 3000
Page 2: ...AS Datascope Anesthesia Delivery System Service Manual...
Page 9: ...Table of Contents vi 0070 10 0683 AS3000 Service Manual This page intentionally left blank...
Page 13: ...Introduction Notes x 0070 10 0683 AS3000 Service Manual This page intentionally left blank...
Page 179: ...0070 10 0683 Rev K August 8 2014...