Installation Checkout Procedure
Installation Guide
2 - 10
9. Test the O
2
:N
2
O ratio system (applicable to -02 units only).
a.
Set all flow control knobs to minimum.
b.
Set the flow of O
2
to 1 L/min.
c.
Open the N
2
O flow knob and verify that the N
2
O flow will not increase higher than
3.7 L/min.
d.
Return flow control knobs to minimum.
10. Perform System Test.
a.
b.
Wait until the System Self Test is complete
c.
Select
Continue
to prompt to the Safety Valve test
FIGURE 2-2
Startup Self Test