1 - 25
Theory of Operation
Ventilator Control and Drive
Inspire Valve, J12
PIN NAME FUNCTION
1 - 2
24V
24V
3 - 4
Valexp inspire
Valve Drive
Man/Auto Valve, J14
PIN NAME FUNCTION
1. 24V
24V
2.
Valman
Manual/Auto Valve Drive
Signals and Power of Sensor Board, J13
PIN NAME FUNCTION
1. A+12V Analogue
+12V
2. FL_INS Import
flow
3. FL_EXP Export
flow
4. FM_FL Flowmeter
flow
5.
Pair
Pressure of air
6.
Paw
Pressure inside air way
7.
Pgas
Pressure gas supply status
8. GND Ground
9. AGND Analog
Ground
10. AGND Analog
Ground
11. P24V
Power
24V
12. P24V
Power
24V
13. P24V
Power
24V
14. P12V
Power
12V
15. P12V
Power
12
16. P5V
Power
5V
17. P5V
Power
5V
18. Inhale Inspire
Valve
19. Man
MANUAL/AUTO
Valve
20. Exhale Expire
Valve
21. Exhale Expire
Valve
22. AGND Analog
Ground
23. AGND Analog
Ground
24. AGND Analog
Ground
25. AGND Analog
Ground