Components
Theory of Operation
1 - 8
26. Vaporizer Storage Mount
The Vaporizer Storage Mount provides the
with auxiliary placement for a
vaporizer.
27. Mains Supply Assembly
The Mains Supply Assembly attaches the
to an external AC wall outlet. It also
supplies power throughout the unit.
28. Electronic Flowmeter Assembly
The Electronic Flowmeter Assembly provides fresh gas flow measurement for the
.
29. Gas Box Assembly
The Gas Box Assembly provides drive gas to the Breathing System and includes a safety
valve to insure that pressure stays below 85 cmH
2
O. It also contains the PEEP valve, and the
differential pressure sensors.
30. YOKE Assembly
The YOKE Assembly provides connection for three backup gas cylinders. The pressure
regulator for the backup gas cylinder reduces the pressure of a high pressure gas cylinder
down to 43.5 psi (300 kPa).
31. AMP Cables
The AMP Cables provide signal transmission between the electric box and gas circuit box.
32. Electric Box Assembly
The Electric Box Assembly provides, and distributes power to the
33. Rear Panel Assembly
The Rear Panel Assembly protects, and provides heat dissipation for the electric box and
auxiliary support for the backup gas cylinder.
1.3.4
Other components (not identified in graphics)
34. Inspiratory and Expiratory Valves
The Inspiratory and Expiratory Valves are unidirectional valves that allow air to flow in only
one direction.
35. Absorber Canister
The
uses two Absorber Canisters which can contain 1500 mL of soda lime each.
They can be used for 6 - 8 hours each if full. Water generated from the reaction with CO
2
is
drained from a valve on the lower side of the canister.
36. APL (Airway Pressure Limiting) Valve
The APL Valve is used for limiting maximum airway pressure during manual ventilation. The
adjustable range of the APL is 0 - 70 cmH
2
O.
Summary of Contents for Datascope AS 3000
Page 2: ...AS Datascope Anesthesia Delivery System Service Manual...
Page 9: ...Table of Contents vi 0070 10 0683 AS3000 Service Manual This page intentionally left blank...
Page 13: ...Introduction Notes x 0070 10 0683 AS3000 Service Manual This page intentionally left blank...
Page 179: ...0070 10 0683 Rev K August 8 2014...