Repair I
nfor
m
ation
Tr
oubleshoo
ting Char
t
In leakage detection, Safety
valve detection failed.
Startup test
1
The proportional valve on the airway
module failed.
2
The safety valve in the airway module
failed.
1
Calibrate the inspiration valve again.
2
Replace the airway module.
In leakage detection, the
leakage volume is 501 -
999mL.
Startup Test
1
The tubing connection is not tight.
2
Incorrect installation of the absorber
canister.
3
The pressure gauge or oxygen sensor on the
circuit is not installed tightly.
4
There is a leakage in the fresh gas supply.
1
Check the connection on the tubing.
2
Check the installation of the absorber canister.
3
Check the installation of the pressure gauge or
oxygen sensor on circuit.
4
Check the fresh gas supply for leakage
In leakage detection, the
leakage volume is larger than
1L.
Startup Test
1
The tubing connection is not tight.
2
Incorrect installation of the absorber
canister.
3
The pressure gauge or oxygen sensor on the
circuit is not installed tightly.
4
There is a leakage in the fresh gas supply.
1
Check the connection on the tubing.
2
Check the installation of the absorber canister.
3
Check the installation of the pressure gauge or
oxygen sensor on circuit.
4
Check the fresh gas supply for leakage
In compliance test,
compliance test failed.
Startup Test
1
The Y-fitting is not sealed or the Breathing
System has leakage.
2
The respiration tubes are too long.
3
Calibration of the pressure sensor is
inaccurate.
4
The flow meter is set incorrectly
5
The calibration of the flow meter is
inaccurate
1
Check the connection and leakage of
respiration tubes.
2
The respiration tubes cannot be too long.
3
Calibrate the pressure oxygen again.
4
Set the flow meter to the correct value
5
Recalibrate the flow meter
The power supply switch on
the main unit is defective.
All
1
The power supply switch on the main unit or
the cable connection failed.
2
Communication board is defective.
1
Check the connection cable of the power
supply switch of the main unit, or replace the
power supply switch of main unit.
2
Replace the communication board.
All or some keys are non
responsive.
All
1
The keyboard is defective.
2
The communication board is defective.
1
Replace the user interface assembly.
2
Replace the user interface assembly.
The encoder is non
responsive.
All
1
The encoder failed.
2
The communication board is defective.
1
Replace the user interface assembly.
2
Check and replace the user interface assembly.
All or some of the work lights
are defective.
All
1
The work light circuit board is defective.
2
The cable is defective
1
Replace the work light circuit board.
2
Check the connection to the power supply.
The switch on the work light
is defective.
All
1
The switch on the work light is defective.
1
Replace the work light switch.
The flow meter light is non
operational.
All
1
The power supply to the light is defective
2
The foil of light is defective.
3
The inverter is defective.
1
Replace the power supply to the light.
2
Replace the foil of light.
3
Replace the inverter
MESSAGE/ISSUE
OCCURRENCE
CAUSE
SOLUTION
Summary of Contents for Datascope AS 3000
Page 2: ...AS Datascope Anesthesia Delivery System Service Manual...
Page 9: ...Table of Contents vi 0070 10 0683 AS3000 Service Manual This page intentionally left blank...
Page 13: ...Introduction Notes x 0070 10 0683 AS3000 Service Manual This page intentionally left blank...
Page 179: ...0070 10 0683 Rev K August 8 2014...