6 - 35
Periodic Maintenance
Functional Tests
6.8.9.4
Auxiliary Flowmeter
1.
Verify an AIR flow of 15 L/min can be obtained by connecting the auxiliary AIR hose to
the pressure source and opening the flow meter.
2.
Verify an O
2
flow of 15 L/min can be obtained by connecting the auxiliary O
2
hose to
the pressure source and opening the flow meter.
6.8.9.5
Patient Suction Regulator (if available)
1.
Set the suction regulator’s selection dial to
LINE
.
2.
Verify maximum suction vacuum on the regulator's gauge.
3.
Set the suction regulator's selection dial to
REGULATE
.
4.
Verify the suction vacuum is adjustable on the regulator's gauge.
5.
Set the suction regulator's selection dial to
OFF
.
6.8.10
Vaporizers
6.8.10.1
Vaporizer Interlock Test
1.
Attach two vaporizers to the Vaporizer Mounting Manifold and lock them in place.
2.
Rotate either of the vaporizer dial to
3%
agent.
3.
Verify that the other vaporizer dial cannot be rotated to a setting.
4.
Set both vaporizer dials to
0
.
5.
Rotate the other vaporizer dial to
3%
.
6.
Verify that the first vaporizer dial cannot be rotated.
7.
Rotate both vaporizer dials to
T
and remove both vaporizers.
8.
Verify that the locking spring is intact.
9.
Reconnect both vaporizers to the Vaporizer Mounting Manifold.
Summary of Contents for Datascope AS 3000
Page 2: ...AS Datascope Anesthesia Delivery System Service Manual...
Page 9: ...Table of Contents vi 0070 10 0683 AS3000 Service Manual This page intentionally left blank...
Page 13: ...Introduction Notes x 0070 10 0683 AS3000 Service Manual This page intentionally left blank...
Page 179: ...0070 10 0683 Rev K August 8 2014...