Installation Checkout Procedure
Installation Guide
2 - 18
NOTE:
If the system is going to be used during the calibration,
insert the O
2
cell plug into the port from which the oxygen
sensor was removed using a push and turn motion.
e.
After at least 3 minutes have passed, select the
Next
button to initiate the
calibration process. The progress bar shown in FIGURE 2-15 or FIGURE 2-17will be
displayed.
FIGURE 2-14
Oxygen Sensor
Calibration Instructions
FIGURE 2-15
Oxygen Sensor Calibration
Progress Bar
FIGURE 2-16
Oxygen Sensor
Calibration Instructions (UI
versions 2.24 and lower)
FIGURE 2-17
Oxygen Sensor Calibration
Progress Bar (UI versions
2.24 and lower)