5 - 7
Calibration
Calibration Procedures
5.5.3
Flow Sensor Calibration
1. Preparing the unit
a.
Remove the Inspiratory and Expiratory valve disks, and reinstall the valve covers.
b.
Connect a Respiration Tube between the Inspiratory and Expiratory Ports. (see
FIGURE 5-2)
c.
Use the
Test Plug to block the Bag Port. (see FIGURE 5-3)
d.
Set the APL Valve to
70
cmH
2
O.
e.
With the Fresh Gas flow set to 100 mL/min, press the O
2
Flush button. Verify that the
Paw gauge reads a minimum of 40 cmH
2
O after releasing the button.
f.
If 40 cmH
2
O cannot be attained, open and close the APL Valve and repeat step
prior to leakage troubleshooting.
g.
Set the Fresh Gas flow to 0 mL/min.
h.
Select
Continue
.
FIGURE 5-2
Connecting the
Respiration Tube
FIGURE 5-3
Blocking the Bag Port