Calibration Procedures
Calibration
5 - 8
2. Calibration
FIGURE 5-4
Connecting the Vent Tester to the
a.
Connect one end of the Calibration Hose to the High Pressure port on the
and the other to the inlet on the Vent Tester.
b.
Connect one end of the Respiration Tube to the expiratory port on the
the other to the outlet of the Vent Tester.
c.
Connect an unterminated tube (at least 1 meter long) to the inspiratory port on the
.
d.
Select
Continue
to start calibration.
e.
Ensure that there is 0 flow on the Flow Analyzer.
f.
Set the on screen calibration to
0
, and select
Done
to begin calibration. Successful
calibration will be confirmed by illuminating the corresponding calibration value
yellow. If the calibration value illuminates red, recalibrate the value until it
successfully calibrated.
g.
Set the onscreen calibration value to
1
.
h.
Raise the flow until it reads 1.00 on the Flow Analyzer using the calibration hose
valve.
i.
Select
Done
to begin calibration.
j.
Repeat steps
g.
through
i.
for all calibration values listed on the screen. Ensure all
successfully calibrate and illuminate yellow.
k.
After all values are successfully calibrated select
Finish Cal
.
l.
The
AS3000
will confirm that calibration was successfully completed.
m.
Select
Done
to return to the calibration screen.
NOTE:
The flow should not be fluctuating in during the calibration.
Summary of Contents for Datascope AS 3000
Page 2: ...AS Datascope Anesthesia Delivery System Service Manual...
Page 9: ...Table of Contents vi 0070 10 0683 AS3000 Service Manual This page intentionally left blank...
Page 13: ...Introduction Notes x 0070 10 0683 AS3000 Service Manual This page intentionally left blank...
Page 179: ...0070 10 0683 Rev K August 8 2014...