6 - 7
Periodic Maintenance
Check Valve Cleaning
3. Remove Breathing System parts:
1.
Remove the following parts from the Breathing System:
• APL Valve
• Breathing Circuit
• Manual Bag
• Bag Arm
• Bellows Cover
• Airway Pressure (PAW) Gauge
• Bellows
• Bellows Cover Base Gasket
• O2 Sensor Assembly
• Inspiratory Check Valve Assy
• Drive Gas Tube
• CGO Hose
• Heater Cable
• Pneumatc Hose
• Scavenger Hose
• Bellows Support Cage (if available)
FIGURE 6-4
Breathing System
Bellows Cover
Expiratory
Check Valve,
Valve Holder,
Cover, and Nut
O2 Sensor
Assembly
APL Valve
Bellows
PAW Gauge
Bag Arm
Inspiratory
Check Valve,
Valve Holder,
Cover, and Nut
Summary of Contents for Datascope AS 3000
Page 2: ...AS Datascope Anesthesia Delivery System Service Manual...
Page 9: ...Table of Contents vi 0070 10 0683 AS3000 Service Manual This page intentionally left blank...
Page 13: ...Introduction Notes x 0070 10 0683 AS3000 Service Manual This page intentionally left blank...
Page 179: ...0070 10 0683 Rev K August 8 2014...