5 - 11
Calibration
Calibration Procedures
2. 30 cmH
2
O Calibration
FIGURE 5-7
Connecting the Y-fitting to the Test Port
a.
Set the APL Valve to
30
cmH
2
O.
b.
Connect the Y-fitting to the Test Port (see FIGURE 5-7)
c.
Connect a Pressure Meter between the Y-fitting and the
(use an adapter if
necessary).
d.
Use the O
2
Flush Valve and O
2
Flow Meter to maintain pressure at 30.0 ±0.2
cmH
2
O on the Pressure Meter.
e.
Select
Continue
to start calibration.
f.
Follow the on-screen instructions.
5.5.6
PEEP Valve
FIGURE 5-8
Disconnect the plastic flaps from the Adult Test Lung
1. Calibration
a.
Disconnect the plastic flaps from the Adult Test Lung (P/N 0138-00-0012) and
connect it to the Y-fitting.
b.
Set the Fresh Gas flow to 0 mL/min.
c.
Press the O2 Flush Valve to fully fill bellows.
d.
Select
Continue
to start calibration.
e.
Follow the on-screen instructions.